Title of article :
Development of amperometric glucose biosensor based on glucose oxidase co-immobilized with multi-walled carbon nanotubes at low potential
Author/Authors :
Rahman، نويسنده , , M.M. and Umar، نويسنده , , Ahmad and Sawada، نويسنده , , Kazuaki، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Abstract :
The electrodes modified with multi-walled carbon nanotube (MWCNT) film containing adsorbed glucose oxidase (GOx) with respect to highly sensitive glucose detection are developed and demonstrated by electrochemical studies at low potential. Glucose-sensing properties were studied using cyclic voltammetric and chronoamperometric techniques. A linear calibration plot was obtained in the concentration range between 1.0 and 500.0 μM, and the detection limit was determined to be 1.3 ± 0.1 μM. Interferences from other biological compounds were studied. The long-term stability, reproducibility, and sensitivity of the GOx biosensor were measured. The proposed glucose biosensor was successfully applied to human serum sample.
Keywords :
MWCNT , Cyclic voltammetry , Glucose oxidase , chronoamperometry
Journal title :
Sensors and Actuators B: Chemical
Journal title :
Sensors and Actuators B: Chemical