• Title of article

    Electromagnetic induction detector with a vertical coil for capillary electrophoresis and microfluidic chip

  • Author/Authors

    Chen، نويسنده , , Zuan-guang and Li، نويسنده , , Ou-lian and Liu، نويسنده , , Cui and Yang، نويسنده , , Xiu-juan، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2009
  • Pages
    4
  • From page
    130
  • To page
    133
  • Abstract
    We report a new electromagnetic induction detector with two vertical inductors for capillary electrophoresis and microfluidic chip capillary electrophoresis. Compared with currently available detection methods, this new detector has several advantages, including simple construction, no complicated elements, ease of assembly and operation, and potential for universal applications. The conditions affecting the response of the detector, including the number of coil turns, the frequency and the effective voltage, were optimized. Its feasibility and performance were demonstrated by analyzing inorganic ions. The new detector showed a well-defined correlation between the ion concentrations and the responses (r = 0.993–0.998), with a detection limit of 5.6 μmol L−1 for Mn2+, as well as good reproducibility and stability for both CE and microfluidic chip CE.
  • Keywords
    Electromagnetic induction , detection , Capillary electrophoresis , Microfluidic chip , Vertical coil
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2009
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number