• Title of article

    A high resolution MEMS based gas chromatography column for the analysis of benzene and toluene gaseous mixtures

  • Author/Authors

    Sun، نويسنده , , Jianhai and Cui، نويسنده , , Dafu and Li، نويسنده , , Yutai and Zhang، نويسنده , , Lulu and Chen، نويسنده , , Jian and Li، نويسنده , , Hui and Chen، نويسنده , , Xing، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2009
  • Pages
    5
  • From page
    431
  • To page
    435
  • Abstract
    his paper reports a high resolution gas chromatography (GC) column based on micro electro mechanical systems (MEMS) technology. This 6.0 m long, 100 μm wide, and 100 μm deep column was fabricated using deep reactive-ion etching (DRIE) to form channels for gas separation, and the channels were sealed with Pyrex 7740 glass by using anode bonding. After the GC column was coated with dimethyl polysiloxane (OV-1) as the stationary phase, benzene and toluene were successfully separated in less than 185 s. The resolution of benzene and toluene was 6.33, which was higher than any previously reported values to our best knowledge. The tailing factors of benzene and toluene were 1.13 and 1.20, respectively, and the number of theoretical plates of toluene was 4850. The system is applicable as a portable device for ambient air quality monitoring and industrial exhaust gas analysis.
  • Keywords
    Gas chromatography column , high resolution , MEMS , DRIE
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2009
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number