• Title of article

    Field-effect calcium sensor for the determination of the risk of urinary stone formation

  • Author/Authors

    Ralph and Beging، نويسنده , , Stefan and Mlynek، نويسنده , , Daniela and Hataihimakul، نويسنده , , Sudkanung and Poghossian، نويسنده , , Arshak and Baldsiefen، نويسنده , , Gerhard and Busch، نويسنده , , Heinz and Laube، نويسنده , , Norbert and Kleinen، نويسنده , , Lisa and Schِning، نويسنده , , Michael J.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    6
  • From page
    374
  • To page
    379
  • Abstract
    Urinary stone formation has been evolved to a widespread disease during the last years. The reason for the formation of urinary stones are little crystals, mostly composed of calcium oxalate, which are formed in human kidneys. The early diagnosis of the risk for urinary stone formation of patients can be determined by the “Bonn-Risk-Index” method based on the potentiometric detection of the Ca2+-ion concentration and an optical determination of the triggered crystallisation of calcium oxalate in unprocessed urine. In this work, miniaturised capacitive field-effect EMIS (electrolyte-membrane-insulator-semiconductor) sensors have been developed for the determination of the Ca2+-ion concentration in human native urine. The Ca2+-sensitive EMIS sensors have been systematically characterised by impedance spectroscopy, capacitance–voltage and constant–capacitance method in terms of sensitivity, signal stability and response time in both CaCl2 solutions and in native urine. The obtained results demonstrate the suitability of EMIS sensors for the measurement of the Ca2+-ion concentration in native urine of patients.
  • Keywords
    Urolithiasis , Bonn-Risk-Index , Field-effect , calcium sensor , EMIS sensor
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2010
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number