Title of article :
A monolithically fabricated flexible resonant circuit for catheter tracking in magnetic resonance imaging
Author/Authors :
M.C. and Ellersiek، نويسنده , , Dennis and Fassbender، نويسنده , , Holger and Bruners، نويسنده , , Philipp and Pfeffer، نويسنده , , Joachim G. and Penzkofer، نويسنده , , Tobias and Mahnken، نويسنده , , Andreas H. and Schmitz-Rode، نويسنده , , Thomas and Mokwa، نويسنده , , Wilfried and Schnakenberg، نويسنده , , Uwe، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2010
Abstract :
To track a flexible catheter with a magnetic resonance imaging (MRI) scanner a wireless resonant circuit has been monolithically integrated into a less than 50 μm thin foil of polyimide. This biocompatible device, protected by a coating of parylene-C, can be attached to catheters with a diameter of at least 5 French. Detailed simulations and measurements have been performed to insure a perfect operation on the instrument for different surrounding liquids. First MRI images in a 1.5 T tomograph show clearly tracking spots with a diameter of about 2.5 mm.
Keywords :
Magnetic resonance imaging (MRI) scanner , Magnetic resonance imaging (MRI) tomography , Biocompatible flexible resonant circuit device , Catheter tracking in magnetic resonance imaging (MRI) , Magnetic Field Simulation , Thin film technology
Journal title :
Sensors and Actuators B: Chemical
Journal title :
Sensors and Actuators B: Chemical