Title of article :
Solution of steady-state substrate concentration in the action of biosensor response at mixed enzyme kinetics
Author/Authors :
Manimozhi، نويسنده , , P. and Subbiah، نويسنده , , A. M. Rajendran، نويسنده , , L.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2010
Abstract :
A mathematical modeling of an amperometric biosensor at mixed enzyme kinetics and diffusion limitation in the case of substrate inhibition is developed. The model is based on non-stationary diffusion equation containing a non-linear term related to non-Michaelis–Menten kinetics of the enzymatic reaction. This paper presents the complex numerical methods (Heʹs variational iteration method and Homotopy perturbation method) to solve the non-linear differential equations that describe the diffusion coupled with a non-linear reaction terms. Approximate analytical expressions for substrate concentration have been derived for all values of parameters α, β and K.
Keywords :
Substrate Inhibition , Non-linear equation , Variational iteration method , Homotopy perturbation method , Amperometric biosensor , reaction diffusion system
Journal title :
Sensors and Actuators B: Chemical
Journal title :