Title of article :
Amperometric detection of phenolic compounds with enzyme immobilized in mesoporous silica prepared by electrophoretic deposition
Author/Authors :
Shimomura، نويسنده , , Takeshi and Itoh، نويسنده , , Tetsuji and Sumiya، نويسنده , , Touru and Hanaoka، نويسنده , , Taka-aki and Mizukami، نويسنده , , Fujio and Ono، نويسنده , , Masatoshi، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Pages :
8
From page :
361
To page :
368
Abstract :
Laccase was immobilized in mesoporous silica powder with a 7.0 nm pore diameter (FSM7.0) coated onto a glassy carbon electrode using an electrophoretic deposition technique, and the electrode was then applied to the amperometric detection of catechol, which is a typical phenolic compound. The behavior of a biosensor attached to the electrode was examined in terms of pH, applied potential, sensitivity and operational range, selectivity, and storage stability. The sensor showed an optimum response at a pH of 5.0 and at an applied potential of −50 mV. The determination range and the response time for catechol were 2.0–100 μM and approximately 2 min, respectively. In addition, the sensor was quite stable and retained its initial response without notable change after being stored for over 50 days. This result suggests that our method is quite useful for the fabrication of a high-performance biosensor for practical use.
Keywords :
Laccase , Mesoporous silica powder , Electrophoretic Deposition , Biosensor , Phenolic compounds
Journal title :
Sensors and Actuators B: Chemical
Serial Year :
2011
Journal title :
Sensors and Actuators B: Chemical
Record number :
1439312
Link To Document :
بازگشت