Title of article :
Sensitive amperometric determination of chemical oxygen demand using Ti/Sb–SnO2/PbO2 composite electrode
Author/Authors :
Ma، نويسنده , , Chuanjun and Tan، نويسنده , , Feng and Zhao، نويسنده , , Huimin and Chen، نويسنده , , Shuo and Quan، نويسنده , , Xie، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Abstract :
A novel Ti/Sb–SnO2/PbO2 composite electrode was fabricated for COD determination. The new electrode configuration improved the sensitivity of the amperometric method apparently. Effects of common experimental parameters, such as applied potential, pH and concentration of the electrolyte on its analytical performance were investigated. A linear range of 0.5–200 mg L−1 COD and a detection limit (a signal-to-noise ratio of 3) of 0.3 mg L−1 were achieved under optimized conditions. The experiments for detecting COD in model samples and real samples were carried out to evaluate the electrodeʹs performance. The obtained results were in good agreement with those determined by the standard dichromate method, with a relative error less than 12%.
Keywords :
Ti/Sb–SnO2/PbO2 electrode , Chemical oxygen demand , amperometric detection
Journal title :
Sensors and Actuators B: Chemical
Journal title :
Sensors and Actuators B: Chemical