Title of article :
On-chip Fabry–Pérot interferometric sensors for micro-gas chromatography detection
Author/Authors :
Reddy، نويسنده , , Karthik and Guo، نويسنده , , Yunbo and Liu، نويسنده , , Jing and Lee، نويسنده , , Wonsuk and Khaing Oo، نويسنده , , Maung Kyaw and Fan، نويسنده , , Xudong، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Pages :
6
From page :
60
To page :
65
Abstract :
We fabricated and characterized on-chip Fabry–Pérot (FP) vapor sensors for the development of on-column micro-gas chromatography (μGC) detectors. The FP sensors were made by coating a thin layer of polymer on a silicon wafer. The air–polymer and polymer–silicon interfaces form an FP cavity, whose resonance wavelengths change in response to the vapor absorption/desorption, thus allowing for rapid detection and quantification of vapors. For proof-of-concept, two polymers (PDMS and SU-8) were used independently and placed in an array in a microfluidic channel, and showed different sensitivities for different vapors. A sub-nano-gram detection limit and sub-second response time were achieved, representing orders of magnitude improvement over those previously reported. This on-chip design will enable the unprecedented integration of optical vapor sensors with μGC systems.
Keywords :
Vapor sensing , Optical vapor sensor , Fabry–Pérot cavity , Micro-gas chromatography , volatile organic compounds
Journal title :
Sensors and Actuators B: Chemical
Serial Year :
2011
Journal title :
Sensors and Actuators B: Chemical
Record number :
1440027
Link To Document :
بازگشت