• Title of article

    Simulation and experimental study of PCR chip based on silicon

  • Author/Authors

    Weiping، نويسنده , , Yan and Liqun، نويسنده , , Du and Jing، نويسنده , , Wang and Lingzhi، نويسنده , , Ma and Jianbo، نويسنده , , Zhu، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    5
  • From page
    695
  • To page
    699
  • Abstract
    PCR technology has been widely applied to biomedicine and other related fields. PCR biochip shortens reaction time and reduces expensive reagents quantities. If the PCR biochip can be integrated with capillary electrophoresis chip, then the microminiaturization of biochemical analysis system will be possible. A micro PCR biochip based on silicon is designed and fabricated by MEMS technology in this paper. In order to reduce the processing complicacy, the temperature sensor and the micro heater were fabricated with chromium metal film. Since the characteristics of metal membrane is closely related to the performance of the material itself and the fabrication technology, the impact of fabrication technology on the membrane performance is studied in this paper. The results indicate the rational domain design, suitable processing condition and appropriate controlling system are preferable to the requirements of PCR reaction.
  • Keywords
    PCR chip , MEMS , heater , Cr metal film , Silicon
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2005
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number

    1443386