Title of article :
Monitoring of ethanol during fermentation using a microbial biosensor with enhanced selectivity
Author/Authors :
Tkac، نويسنده , , Jan and Vostiar، نويسنده , , Igor and Gemeiner، نويسنده , , Peter and Sturdik، نويسنده , , Ernest، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Pages :
3
From page :
127
To page :
129
Abstract :
The present study is concerning the construction of ferricyanide-mediated Gluconobacter oxydans cell ethanol biosensor. The size exclusion effect of a cellulose acetate membrane was used for elimination of glucose interferences during ethanol assays in real samples. A typical response time of the biosensor was 13 s with a high sensitivity of 3.5 μA mM−1. The microbial biosensor exhibits a very low detection limit of 0.85 μM and a wide linear range from 2 to 270 μM. The operational stability was excellent. During 8.5 h of repetitive ethanol assays, no decrease in the sensor sensitivity was observed. The biosensor was successfully used in the off-line monitoring of ethanol fermentation with a good agreement with HPLC measurements (R2=0.998).
Keywords :
Ethanol fermentation , Cellulose acetate membrane , Amperometric biosensor , Gluconobacter oxydans