Title of article
Determination of malondialdehyde in breath condensate by high-performance liquid chromatography with fluorescence detection
Author/Authors
Lنrstad، نويسنده , , Monica and Ljungkvist، نويسنده , , Gِran and Olin، نويسنده , , Anna-Carin and Torén، نويسنده , , Kjell، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2002
Pages
8
From page
107
To page
114
Abstract
An automated and rapid method for quantifying malondialdehyde (MDA) in breath condensate was developed and validated. The method is based on derivatisation with thiobarbituric acid, HPLC separation and fluorescence detection and is optimised for determination of MDA in breath condensate. Sample collection is non-invasive and simple. The detection limit (4.1 nM) is low, precision is good and the analysis time is short. The response is linear in the concentration range of 0.020 to 1.0 μM. Samples could be stored for 1 month at −20°C and for 3 months at −80°C without losses. Using this method, there was no statistically significant difference between patients with asthma and patients without asthma. However, among females, subjects with asthma had higher MDA levels as compared to females without asthma (0.17 vs. 0.12 pmol/s, p=0.04). The use of the method when studying airway inflammation has to be further evaluated.
Keywords
Malondialdehyde
Journal title
Journal of Chromatography B
Serial Year
2002
Journal title
Journal of Chromatography B
Record number
1453006
Link To Document