• Title of article

    Rapid and sensitive method for the analysis of carbon monoxide in blood using gas chromatography with flame ionisation detection

  • Author/Authors

    Sundin، نويسنده , , Anna-Maja and Larsson، نويسنده , , Jan Erik، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2002
  • Pages
    7
  • From page
    115
  • To page
    121
  • Abstract
    In order to measure changes in physiological CO concentrations in blood with good accuracy, a method was developed using gas chromatography with flame ionisation detection (250°C). A nickel catalyst system was fitted to convert CO to methane at 375°C after separation with a molecular sieve column at 35°C. Helium was used as carrier at 30 ml/min. Porcine or human blood (400 μl) was sampled in gastight tubes and treated with sulfuric acid and saponin (800 μl). Accuracy was 1.4% and 1.5% (RSD), respectively. Precision was 2.8% (porcine blood). Limit of detection was 0.01 nmol/ml gas and limit of quantification 12 nmol/ml blood. Calibration was made in the interval 12–514 nmol/ml blood (corresponding to 0.1–6% COHb). Samples were stable for at least a month at +4°C. This paper describes a method with high sensitivity and good accuracy, suitable for analysis of low CO concentrations.
  • Keywords
    CARBON MONOXIDE
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2002
  • Journal title
    Journal of Chromatography B
  • Record number

    1453007