Title of article :
Sensitive liquid chromatography–tandem mass spectrometry method for the determination of clarithromycin in human plasma
Author/Authors :
van Rooyen، نويسنده , , G.F and Smit، نويسنده , , M.J and de Jager، نويسنده , , A.D and Hundt، نويسنده , , H.K.L and Swart، نويسنده , , K.J and Hundt، نويسنده , , A.F، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Pages :
7
From page :
223
To page :
229
Abstract :
A sensitive method for the determination of clarithromycin in plasma is described, using high-performance liquid chromatographic separation with tandem mass spectrometric detection. Samples were prepared using liquid–liquid extraction and separated on a Supelco Discovery® C18 column with a mobile phase consisting of acetonitrile, methanol and acetic acid. Detection was performed by a PE SCIEX API 2000 mass spectrometer in the multiple reaction monitoring (MRM) mode (LC–MS–MS) using TurboIonSpray ionization and monitoring the transition of the protonated molecular ion for clarithromycin at m/z 748.5 (M+1) to the predominant product ion of m/z 158.2. The mean recovery of clarithromycin was 87.3%, with a lower limit of quantification of 2.95 ng/ml when using 0.3-ml plasma. This high-throughput method was used to quantify 230 samples per day, and is sufficiently sensitive to be employed in pharmacokinetic studies.
Keywords :
Clarithromycin
Journal title :
Journal of Chromatography B
Serial Year :
2002
Journal title :
Journal of Chromatography B
Record number :
1453244
Link To Document :
بازگشت