Title of article
Determination of 235U and 238U in urine samples using sector field inductively coupled plasma mass spectrometry
Author/Authors
Schramel، نويسنده , , P.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2002
Pages
4
From page
275
To page
278
Abstract
The high sensitivity of SF-ICP-MS (sector field inductively coupled plasma mass spectrometry) using a torch with the “guard-electrode” (capacitive decoupled plasma) allows the determination of 238U (isotope abundance 99.2%) and 235U (0.8%) and their isotope ratio in human urine samples down to the physiological level of <10 ng/l total uranium. For sample preparation UV photolysis was used. Some quality criteria like for the detection limit, the reproducibility, recovery and the isotope ratio are given. The method can be applied in occupational as well as in environmental medicine because of its outstanding detection power.
Keywords
uranium
Journal title
Journal of Chromatography B
Serial Year
2002
Journal title
Journal of Chromatography B
Record number
1454131