• Title of article

    Development and validation of a liquid chromatographic method for in vitro mupirocin quantification in both skin layers and percutaneous penetration studies

  • Author/Authors

    Echevarr??a، نويسنده , , Lydia and Blanco-Pr??eto، نويسنده , , Mar??a J and Campanero، نويسنده , , Miguel Angel and Santoyo، نويسنده , , Susana and Ygartua، نويسنده , , Pilar، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    9
  • From page
    233
  • To page
    241
  • Abstract
    A simple, rapid, and sensitive reversed-phase high-performance liquid chromatographic (HPLC) method for the measurement of mupirocin concentrations in both skin layers and percutaneous samples has been developed. Mupirocin was extracted from skin layers using PBS–acetonitrile (90:10, v/v). The method is sufficiently sensitive and repeatable to be used in percutaneous penetration studies. The samples were chromatographed on a 250 mm×4 mm C8 LiChrospher Select B (5 μm). The mobile phase composition was a mixture of acetonitrile–ammonium acetate 0.05 M (27.5:72.5, v/v) adjusted to pH 6.3 with acetic acid. The analyte was detected at 228 nm and the run time was 11 min. Linearity was confirmed in the concentration range 0.2–20 μg/ml and the limit of detection was 9.5 ng/ml.
  • Keywords
    Skin recovery , Mupirocin
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2003
  • Journal title
    Journal of Chromatography B
  • Record number