Title of article :
Measurement of carvedilol in plasma by high-performance liquid chromatography with electrochemical detection
Author/Authors :
Machida، نويسنده , , Maiko and Watanabe، نويسنده , , Masato and Takechi، نويسنده , , Shigeru and Kakinoki، نويسنده , , Shigeo and Nomura، نويسنده , , Akikazu، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
5
From page :
187
To page :
191
Abstract :
Carvedilol is a β/α1-adrenoceptor blocker. A sensitive method for measuring plasma levels of carvedilol in human administrated low doses is needed since its plasma concentration is low. We measured carvedilol and carvedilol M21-aglycon using high-performance liquid chromatography (HPLC) with electrochmical detection. The amperometric detector was operated at 930 mV versus Ag/AgCl. Mean coefficients of variation (n=5) for carvedilol and M21-aglycon were 4.0 and 7.7% (intra) and 6.1 and 6.7% (inter), respectively. The lower limit of quantification for each analyte was 0.10 ng/ml (signal-to-noise ratio=3). This lower limit of quantification for carvedilol was sufficient for clinical use.