Title of article :
Determination of Tirofiban in human serum by liquid chromatography–tandem mass spectrometry
Author/Authors :
Oertel، نويسنده , , R and Kِhler، نويسنده , , A and Koster، نويسنده , , A and Kirch، نويسنده , , W، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Abstract :
A liquid chromatography–tandem mass spectrometric (LC–MS–MS) method with a rapid and simple sample preparation was developed and validated for the determination of Tirofiban in biological fluids. Tirofiban in serum samples was extracted and cleaned up by using an automated solid phase extraction method. An external calibration was used. The mass spectrometer was operated in the multiple reaction monitoring mode (MRM). A good linear response over the range of 2–200 ng/ml was demonstrated. The accuracy for Tirofiban ranged from 94.8 to 110.8% within-day and from 103.0 to 104.7% between-day. The lower limit of quantification was 2 ng/ml. This method is suitable for pharmacokinetic studies.
Journal title :
Journal of Chromatography B
Journal title :
Journal of Chromatography B