• Title of article

    Development and validation of an HPLC-UV method for determination of iohexol in human plasma

  • Author/Authors

    Soman، نويسنده , , Rohit S. and Zahir، نويسنده , , Hamim and Akhlaghi، نويسنده , , Fatemeh، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    5
  • From page
    339
  • To page
    343
  • Abstract
    An HPLC-UV analytical method for estimation of iohexol in human plasma was developed and validated. Protein precipitation and iohexol extraction from plasma (100 μl) was carried out by adding 800 μl perchloric acid (5%, v/v in water) containing iohexol related compound B as the internal standard followed by vortex mixing and centrifugation. The supernatant (90 μl) was then injected onto a μBondapak C18 column (150 mm × 3.9 mm, 10 μm) maintained at 30 °C. The mobile phase comprised of various proportions of acetonitrile and water with a total run time of 12 min and the wavelength of the UV detector was set at 254 nm. The extraction recovery of iohexol from plasma was >95% and the calibration curve was linear (r2 = 0.99) over iohexol concentrations ranging from 10 to 750 μg/ml (n = 8). The method had an accuracy of >92% and intra- and inter-day CV of <3.7% and <3.6%, respectively. The method reported is simple, reliable, precise, accurate and has the capability of being used for determination of iohexol in clinical settings.
  • Keywords
    GFR , Iohexol
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2005
  • Journal title
    Journal of Chromatography B
  • Record number

    1457237