Title of article :
Development and validation of an LC–MS/MS method for the quantitative determination of aripiprazole and its main metabolite, OPC-14857, in human plasma
Author/Authors :
Kubo، نويسنده , , Masanori and Mizooku، نويسنده , , Yasuo and Hirao، نويسنده , , Yukihiro and Osumi، نويسنده , , Takahiko، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Pages :
6
From page :
294
To page :
299
Abstract :
An accurate, sensitive, reproducible, and selective liquid chromatography/tandem mass spectrometry (LC–MS/MS) method for determination of aripiprazole and its main metabolite, OPC-14857, in human plasma was developed and validated. Chromatographic separation was achieved isocratically on a C18 reversed-phase column within 7.5 min. The calibration curve, ranging from 0.1 to 100 ng/ml, was fitted to a 1/y2-weighted linear regression model. The assay showed no significant interference. Lower limit of quantitation (LLOQ) for both analytes was 0.1 ng/ml using 0.4 ml of plasma. Intra- and inter-assay precision and accuracy values for aripiprazole and OPC-14857 were within regulatory limits.
Keywords :
OPC-14857 , LC–MS/MS , Aripiprazole
Journal title :
Journal of Chromatography B
Serial Year :
2005
Journal title :
Journal of Chromatography B
Record number :
1459141
Link To Document :
بازگشت