    Design of experiments for a precise estimation of the calibration curve of periodate in its determination by flow-injection analysis and chemiluminescence detection

  • Author/Authors

    Papamichael، نويسنده , , Emmanuel M. and Evmiridis، نويسنده , , Nickolaos P. and Thanasoulias، نويسنده , , Nickolaos and Stefanou، نويسنده , , Despina، نويسنده ,

  • Issue Information
    دوفصلنامه با شماره پیاپی سال 1995
  • Pages
    11
  • From page
    227
  • To page
    237
  • Abstract
    A design of experiments method based on the criterion of D-optimality is presented. It is applied to the estimation procedure of the parameters of a periodate calibration curve described by a model equation of the form y = ax2(bx + cx + 1) and giving the maximum possible precision using three experimental measurements. The use of four or five experimental measurements for producing the calibration curve was not found more advantageous than that of three. Simulated data were generated, for the three optimal points, and a large number of parameter estimates were obtained under constant and/or relative errors (in y) showing a robust behaviour. The %-bias and the %-excess variance based on the simulated data were found to be non-significant.
  • Keywords
    Curve fitting , Calibration , Periodate , Chemiluminescence , Pyrogallol , Data processing , Experimental design