ATR-FTIR in monitoring of crystallization processes: comparison of indirect and direct OSC methods
Author/Authors
Pِllنnen، نويسنده , , Kati and Hنkkinen، نويسنده , , Antti and Reinikainen، نويسنده , , Satu-Pia and Louhi-Kultanen، نويسنده , , Marjatta and Nystrِm، نويسنده , , Lars، نويسنده ,
Issue Information
دوفصلنامه با شماره پیاپی سال 2005
Pages
11
From page
25
To page
35
Abstract
A stable PLS calibration for solute concentration measurement in batch cooling crystallization processes is presented. Two example cases are included: Sulfathiazole dissolved in five different mixtures of water and 1-propanol and C15 dissolved in toluene. The data to be used in calibration are validated using Multivariate Statistical Process Control (MSPC) and sensitivity analyses criteria. Two Orthogonal Signal Correction (OSC) filters are compared as preprocessing techniques to enhance the model performance. The inherent criteria for selecting the best performing model was RMSEV of the test set. Models are validated with two different external test sets. One measured in conditions similar to calibration set and another measured in conditions close to true crystallization measurement conditions. The latter one tests the flexibility of the model to be used in future predictions. Results show that careful data and model validation steps clearly improve the modeling procedure. OSC filtering clearly simplified the model and also enhanced the performance of the PLS model. The models have successfully been used in several crystallization experiments.
Keywords
OSC , Multivariate calibration , ATR-FTIR , Cooling crystallization
Journal title
Chemometrics and Intelligent Laboratory Systems
Serial Year
2005
Journal title
Chemometrics and Intelligent Laboratory Systems
Record number
1461408
Link To Document