Title of article :
High-performance liquid chromatography method for the simultaneous determination of sulfamethoxazole and trimethoprim in bovine milk using an on-line clean-up column
Author/Authors :
Pereira، نويسنده , , A.V. and Cass، نويسنده , , Q.B.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Pages :
8
From page :
139
To page :
146
Abstract :
A bidimensional HPLC method for the simultaneous determination of sulfamethoxazole (SMX) and trimethoprim (TMP) in bovine milk has been developed and validated. After centrifugation, aliquots (150 μl) of milk samples were directly injected to a column-switching HPLC system. At the first step a RAM octyl-BSA column was employed to automatically remove proteins that otherwise would interfere with milk analysis. The mobile phase 0.01 M phosphate buffer pH 6.0:acetonitrile (95:5, v/v) was used in the first 5 min for the elution of milk proteins and then 0.01 M phosphate buffer pH 6.0:acetonitrile (83:17, v/v) for transfer SMX and TMP to the analytical column. The separation of SMX and TMP from one another and from other remaining milk components was performed on an octyl column using the mobile phase 0.01 M phosphate buffer pH 5.0:acetonitrile (82:18, v/v), which were detected by UV at 265 nm. The calibration graphs were linear in the concentration ranges of 25–800 ng/ml and 50–400 ng/ml for SMX and TMP, respectively. The intra- and inter-assay coefficients of variation were less than 15% for both drugs. The validated method was applied to the analysis of milk samples of twelve (two groups of six) cows after administration (intramuscular or subcutaneous) of a single recommended therapeutic dose of the SMX–TMP combination.
Keywords :
Restricted access media (RAM) , Column-switching , sulfamethoxazole , Sulfonamide , direct injection , Trimethoprim , Bovine milk
Journal title :
Journal of Chromatography B
Serial Year :
2005
Journal title :
Journal of Chromatography B
Record number :
1462310
Link To Document :
بازگشت