Title of article :
Monolithic column-based reversed-phase liquid chromatography separation for amino acid assay in microdialysates and cerebral spinal fluid
Author/Authors :
Devall، نويسنده , , A.J. and Blake، نويسنده , , R. and Langman، نويسنده , , N. and Smith، نويسنده , , C.G.S. and Richards، نويسنده , , D.A. and Whitehead، نويسنده , , K.J.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
6
From page :
323
To page :
328
Abstract :
The development of a HPLC method using a monolithic C18 column is described using fluorescence detection for the assay of 21 amino acids and related substances with derivatisation using ortho-phthaldialdehyde (OPA) in the presence of 3-mercaptopropionic acid (3-MPA). The method employs a tertiary gradient and has a run time of 24 min. Linearity (r2) for each amino acid was found to be greater than 0.99 up to a 10 μM concentration; reproducibility across all analyses (relative standard deviation (R.S.D.)) was between 0.97 and 6.7% and limit of detection (LOD) between 30 and 300 fmol on column. This method has been applied to the analysis of amino acids in both spinal microdialysis and cerebral spinal fluid samples.
Keywords :
Reversed-phase liquid chromatography , Monolithic , Ortho-phthaldialdehyde , Amino acid
Journal title :
Journal of Chromatography B
Serial Year :
2007
Journal title :
Journal of Chromatography B
Record number :
1464077
Link To Document :
بازگشت