Title of article
Partial within-animal calibration: A new calibration approach in lead optimisation high-throughput bioanalysis
Author/Authors
Dewé، نويسنده , , Walthère and De Smet، نويسنده , , Michaël and Evrard، نويسنده , , Nathalie and Culot، نويسنده , , Benoit and Lastelle، نويسنده , , Marc and Saucez، نويسنده , , Gaëlle and Ingels، نويسنده , , Dominique and Taillieu، نويسنده , , Xavier، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2007
Pages
9
From page
183
To page
191
Abstract
The lead optimization phase of drug discovery requires high-throughput analyses for quantification in biological matrices and in plasma in particular. Over the last decade, some technical innovations allowed the pharmaceutical industry to improve the quality of the results. However, there is room for improvement. In this context, a new calibration strategy is proposed in this paper. Experiments were performed on dog plasma samples and it was showed that a within-animal calibration strategy can reduce the bias up to 20% and improve the precision up to 20%. However, a partial within-animal calibration is preferred to the full approach in order to avoid to many sample preparations.
Keywords
High-Throughput , bias , accuracy , Calibration , Precision , bioanalysis , Validation