• Title of article

    Determination of 5-aminoimidazole-4-carboxamide in human plasma by ion-pair extraction and LC–MS/MS

  • Author/Authors

    Chen، نويسنده , , Xijing and Jing، نويسنده , , Jing and Ren، نويسنده , , Weichao and Han، نويسنده , , De-en and Jing، نويسنده , , Ning and Wang، نويسنده , , Guangji، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    5
  • From page
    140
  • To page
    144
  • Abstract
    A liquid chromatography/tandem mass spectrometry (LC–MS/MS) method was established for the determination of 5-aminoimidazole-4-carboxamide (AICA) in human plasma. The method included a solvent extraction of AICA as an ion pair with 1-pentanesulfonate ion and a separation on a Hypersil ODS2 column with the mobile phase of methanol–water (68:32, v/v). Determination was performed using an electrospray ionization source in positive ion mode (ESI+). Multiple reaction monitoring (MRM) was utilized for the detection monitoring m/z at 127 → 110 for AICA, and 172 → 128 for IS. The calibration curve was linear within a range from 20 to 2000 ng/mL and the limit of quantity for AICA in plasma was 20 ng/mL. RSD of intra-assay and inter-assay were no more than 5.90% and 5.65%.
  • Keywords
    Orazamide , AICA , Ion-pair extraction , LC–MS/MS
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2007
  • Journal title
    Journal of Chromatography B
  • Record number