Determination of 8-iso-prostaglandin F2α in exhaled breath condensate using combination of immunoseparation and LC–ESI-MS/MS
Author/Authors
Syslov?، نويسنده , , Kamila and Ka?er، نويسنده , , Petr and Kuzma، نويسنده , , Marek and Klus??kov?، نويسنده , , Pavl?na and Fenclov?، نويسنده , , Zdena and Lebedov?، نويسنده , , Jind?i?ka and Pelclov?، نويسنده , , Daniela، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2008
Pages
7
From page
8
To page
14
Abstract
Rapid and precise method for the determination of 8-iso-prostaglandin F2α, an essential marker of the oxidative stress, in exhaled breath condensate (EBC) was developed. The protocol consisted of stable isotope dilution, immunoseparation combined with selective and sensitive LC–ESI-MS/MS operated in multiple reaction monitoring (MRM) mode. The imprecision of the developed method was below 8.8%, the parameter of mean inaccuracy was determined as <9.6% (0–250 pg of 8-iso-prostaglandin F2α/ml EBC). The limit of detection (LOD) was 1 pg/ml EBC and limit of quantification (LOQ) 5 pg/ml EBC. A significant difference in 8-iso-prostaglandin F2α content between the group of asbestosis patients and healthy volunteers was found.
Keywords
8-iso-Prostaglandin F2? , Exhaled breath condensate , Immunoseparation
Journal title
Journal of Chromatography B
Serial Year
2008
Journal title
Journal of Chromatography B