• Title of article

    A sensitive HPLC–MS–MS method for the determination of raltegravir in human plasma

  • Author/Authors

    Long، نويسنده , , Mary C. and Bennetto-Hood، نويسنده , , Chantelle and Acosta، نويسنده , , Edward P.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    7
  • From page
    165
  • To page
    171
  • Abstract
    This work describes an assay system that has been developed to quantify raltegravir concentrations in human plasma using a liquid–liquid extraction technique paired with HPLC separation and MS–MS detection. The dynamic range of this assay extends from 1 to 3000 ng/mL, with a coefficient of determination (r2, mean ± SD) of 0.9992 ± 0.0002. The mean precision values for calibration standards ranged from 0.6% to 3.0%, while accuracy values were 96.5–104.3%. This procedure is an accurate, precise, and sensitive method for raltegravir quantitation and was successfully validated using external proficiency testing.
  • Keywords
    HIV , Integrase inhibitor , Raltegravir , human , mass spectrometry
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2008
  • Journal title
    Journal of Chromatography B
  • Record number

    1465925