• Title of article

    Determination of methimazole in urine with the iodine-azide detection system following its separation by reversed-phase high-performance liquid chromatography

  • Author/Authors

    Zakrzewski، نويسنده , , Robert، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    8
  • From page
    67
  • To page
    74
  • Abstract
    The iodine-azide detection system to determine methimazole following its separation by RP-HPLC is described in this paper. The reaction between iodine and azide ions induced by methimazole was applied as a post-column reaction detection system. Neither extraction nor preconcentration of the sample was necessary. The methimazole standards added to normal urine show that the response of the detector, set at 350 nm (corresponding to unreacted iodine in the post-column iodine-azide reaction), was linear within the concentration range 2–10 nmol/mL of urine. The relative standard deviation values for precision and recovery within the calibration range were from 0.3 to 3.2% and from 97 to 102%, respectively. Limits of detection (LOD) and quantitation (LOQ) were 1 and 2 nmol/mL of urine, respectively. The method was applied to the separation and determination of patient urine samples and the analytical results were satisfactory.
  • Keywords
    Iodine-azide reaction , Methimazole assay , urine , HPLC
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2008
  • Journal title
    Journal of Chromatography B
  • Record number