• Title of article

    Validated LC–MS/MS methods for the determination of risperidone and the enantiomers of 9-hydroxyrisperidone in human plasma and urine

  • Author/Authors

    Frederik De Meulder، نويسنده , , Marc and Remmerie، نويسنده , , Bart M.M. and de Vries، نويسنده , , Ronald and Sips، نويسنده , , Luc L.A. and Boom، نويسنده , , Sandra and Hooijschuur، نويسنده , , Edwin W.J. and van de Merbel، نويسنده , , Nico C. and Timmerman، نويسنده , , Philip M.M.B.L.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    9
  • From page
    8
  • To page
    16
  • Abstract
    Two liquid chromatography coupled to tandem mass spectrometry (LC–MS/MS) methods are described, one for the quantitative determination of risperidone and the enantiomers of its active metabolite 9-hydroxyrisperidone (paliperidone) in human plasma and the other for the determination of the enantiomers of 9-hydroxyrisperidone in human urine. The plasma method is based on solid-phase extraction of 200 μl of sample on a mixed-mode sorbent, followed by separation on a cellulose-based LC column with a 13.5-min mobile phase gradient of hexane, isopropanol and ethanol. After post-column addition of 10 mM ammonium acetate in ethanol/water, detection takes place by ion-spray tandem mass spectrometry in the positive ion mode. Method validation results show that the method is sufficiently selective towards the enantiomers of 7-hydroxyrisperidone and capable of quantifying the analytes with good precision and accuracy in the concentration range of 0.2–100 ng/ml. An accelerated (run time of 4.3 min) and equally valid method for the enantiomers of 9-hydroxyrisperidone alone in plasma is obtained by increasing the mobile phase flow-rate from 1.0 to 2.0 ml/min and slightly adapting the gradient conditions. The urine method is based on the same solid-phase extraction and chromatographic approach as the accelerated plasma method. Using 100 μl of sample, (+)- and (−)-9-hydroxyrisperidone can be quantified in the concentration range 1–2000 ng/ml. The accelerated method for plasma and the method for urine can be used only when paliperidone is administered instead of risperidone, as there is insufficient separation of the 9-hydroxy enantiomers from the 7-hydroxy enantiomers, the latter ones being present only after risperidone administration.
  • Keywords
    Risperidone , Paliperidone , Chiral separation
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2008
  • Journal title
    Journal of Chromatography B
  • Record number