Abstract :
Due to its increased concentration in blood, 3-hydroxyisovalerylcarnitine (C5OH-I) is an important indicator for the diagnosis of organic acidemias in newborns. However, C5OH-I has not been used as a standard in tandem mass spectrometric (MS/MS) assays because its isolation is difficult. We developed a new synthesis of C5OH-I and investigated its behavior by MS/MS. A method using the multiple reaction monitoring (MRM) mode of MS/MS with HPLC was developed which provides high accuracy, precision and reproducibility. Acylcarnitine profiles in the serum and urine of a patient with multiple carboxylase deficiency (MCD) showed increased levels compared to a healthy patient.
Keywords :
Acylcarnitine , 3-Hydroxyisovalerylcarnitine , 3-Methylcrotonylcarnitine , Multiple carboxylase deficiency , Holocarboxylase synthetase deficiency , Newborn screening , Tandem mass spectrometry , LC–MS/MS , Tiglylcarnitine