• Title of article

    Liquid chromatography–mass spectrometric assay for the quantitation in human plasma of ABT-888, an orally available, small molecule inhibitor of poly(ADP-ribose) polymerase

  • Author/Authors

    Parise، نويسنده , , Robert A. and Shawaqfeh، نويسنده , , Mohammad and Egorin، نويسنده , , Merrill J. and Beumer، نويسنده , , Jan H.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    7
  • From page
    141
  • To page
    147
  • Abstract
    ABT-888, a poly(ADP-ribose) polymerase (PARP) -inhibitor in clinical trials, potentiates DNA-damaging agents. We developed and validated, according to FDA guidelines, an LC–MS assay for sensitive, accurate and precise quantitation of ABT-888 and its metabolite M8 in 0.2 mL human plasma. After ethyl acetate extraction, separation is achieved with a hydro-Synergi column and a 0.1% formic acid in acetonitrile/water-gradient. Detection uses electrospray, positive-mode ionization mass spectrometry. Between 10 (LOQ) and 1000 ng/mL, accuracy was 95.5–98.5% for ABT-888 and 91.4–100.9% for M8, and precision was 0.1–4.9% for ABT-888 and 0–13.7% for M8. The assay is being applied to samples generated in several clinical trials.
  • Keywords
    ABT-888 , mass spectrometry , PARP-inhibitor
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2008
  • Journal title
    Journal of Chromatography B
  • Record number

    1466301