Optimized method for the determination of itopride in human plasma by high-performance liquid chromatography with fluorimetric detection
Author/Authors :
Pt??ek، نويسنده , , Pavel and Kl?ma، نويسنده , , Josef and Macek، نويسنده , , Jan، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Abstract :
A high-performance liquid chromatographic method with fluorescence detection for the determination of itopride in human plasma is reported. The sample preparation was based on liquid–liquid extraction of itopride from plasma with t-butylmethylether and dichloromethane (70:30, v/v) mixture followed by a back extraction of the analyte to the phosphate buffer (pH 3.2). Liquid chromatography was performed on an octadecylsilica column (55 mm × 4 mm, 3 μm particles), the mobile phase consisted of acetonitrile–triethylamine–15 mM dihydrogenpotassium phosphate (14.5:0.5:85, v/v/v), pH of the mobile phase was adjusted to 4.8. The run time was 3 min. The fluorimetric detector was operated at 250/342 nm (excitation/emission wavelength). Naratriptan was used as the internal standard. The limit of quantitation was 9.5 ng/ml using 0.5 ml of plasma. The method precision and inaccuracy were less than 8%. The assay was applied to the analysis of samples from a bioequivalence study.
Keywords :
Itopride , extraction , HPLC , fluorescence
Journal title :
Journal of Chromatography B
Journal title :
Journal of Chromatography B