Title of article :
Evaluation of uncertainty of measurement from method validation data: An application to the simultaneous determination of retinol and α-tocopherol in human serum by HPLC
Author/Authors :
Semeraro، نويسنده , , Antonella and Altieri، نويسنده , , Ilaria and Patriarca، نويسنده , , Marina and Menditto، نويسنده , , Antonio، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Abstract :
An isocratic RP-HPLC method for the determination of retinol and α-tocopherol in serum, with fluorescence and UV/VIS detection, respectively, was developed and validated according to international guidelines. Detection (retinol 0.015 mg/L, α-tocopherol 0.29 mg/L) and quantification (retinol 0.05 mg/L, α-tocopherol 0.95 mg/L) limits were determined. Repeatability was <3.3% and <2.9% and intermediate precision was <4.6% and <3.2%, for retinol and α-tocopherol, respectively. Certified reference materials were utilised to assess bias and guarantee traceability to SI units. Expanded uncertainties (retinol 8.9%; α-tocopherol 7.9%), estimated according to the EURACHEM/CITAC guide from method validation data, satisfied fit-for-purpose requirements based on biological variability.
Keywords :
serum , ?-Tocopherol , HPLC , Validation , Biological variability , Uncertainty of measurement , Quality specifications , Retinol
Journal title :
Journal of Chromatography B
Journal title :
Journal of Chromatography B