Title of article :
Development and validation of an improved liquid chromatography–mass spectrometry method for the determination of pemetrexed in human plasma
Author/Authors :
Bobin-Dubigeon، نويسنده , , C. and Amiand، نويسنده , , M.B. and Herrenknecht، نويسنده , , C. and Bard، نويسنده , , J.M.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Pages :
6
From page :
2451
To page :
2456
Abstract :
An improved liquid chromatography–mass spectrometry method for the determination of pemetrexed in human plasma was developed and validated using a simple quadrupole LC–MS and a new SPE cartridge (Plexa® Bond Elut). The analysis was achieved with a C18 analytical column using a mobile phase consisting of formic acid/acetonitrile and isocratic flow for 7 min. The linear ranges (r2 > 0.99) were found from 5 to 5000 ng/mL. The lower limit of detection was 2.5 ng/mL. Within-day and between-day precisions were less than 7.2% and inaccuracy did not exceed 2.8%. This new method is suitable to support pharmacokinetic studies and drug monitoring.
Keywords :
Pemetrexed , PLASMA , Pharmacokinetic studies , Liquid chromatography/mass spectrometry