• Title of article

    Determination of methocarbamol concentration in human plasma by high performance liquid chromatography–tandem mass spectrometry

  • Author/Authors

    Zha، نويسنده , , Wuyi and Zhu، نويسنده , , Zhimeng، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    5
  • From page
    831
  • To page
    835
  • Abstract
    A simple and reproducible high performance liquid chromatography–tandem mass spectrometric method was developed for methocarbamol analysis in human plasma. Methocarbamol and the internal standard (IS) were extracted by a protein precipitation method. Under isocratic separation condition the chromatographic run time was 3.0 min. The calibration curve was linear over a range of 150–12,000 ng/mL with good intraday assay and interday assay precision (CV% < 10.9%). The method was proven to be sensitive and selective for the analysis of methocarbamol in human plasma for bioequivalence study.
  • Keywords
    bioequivalence , Methocarbamol , LC/MS/MS , Human plasma
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2010
  • Journal title
    Journal of Chromatography B
  • Record number