Title of article :
Measurement of 25-OH-vitamin D in human serum using liquid chromatography tandem-mass spectrometry with comparison to radioimmunoassay and automated immunoassay
Author/Authors :
van den Ouweland، نويسنده , , Johannes M.W. and Beijers، نويسنده , , Antonius M. and Demacker، نويسنده , , Pierre N.M. and van Daal، نويسنده , , Henny، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2010
Abstract :
The plasma 25-OH vitamin D concentration is a reliable biomarker for vitamin D status but assayʹs variability makes adequate monitoring of vitamin D status difficult. We employed isotope-dilution liquid chromatography (LC) tandem-mass spectrometry (MS/MS) for the measurement of both 25-OH vitamin D3 and 25-OH vitamin D2 in human serum. Hexadeuterium labelled 25-OH vitamin D3 internal standard (IS) was added to calibrators (prepared in phosphate-buffered saline with 60 g/L albumin), controls or patient sera and 25-OH vitamin D metabolites were released from vitamin D binding protein by adding sodium hydroxide prior to protein precipitation by acetonitrile/methanol (9:1, v/v). Subsequent off-line solid-phase extraction was followed by chromatographic separation on a C-18 column using a water/methanol/ammonium acetate gradient. Detection was by Atmospheric Pressure Electrospray Ionisation (AP-EI) followed by selected reaction monitoring. We compared the LC-MS/MS assay to the DiaSorin radioimmunoassay (RIA) and a recently re-standardised version of an automated electrochemiluminescent immunoassay (ECLIA) from Roche Diagnostics. We also analysed external quality control samples from the International Vitamin D External Quality Assessment Scheme (DEQAS) for comparison with other participating laboratories using LC-MS. The method was linear from 5 to at least 550 nmol/L with intra- and interday CVʹs ≤6% for both 25-OH vitamin D3 and 25-OH vitamin D2. Recoveries ranged between 94.9 and 106.9% for 25-OH vitamin D3 and 82.7 and 100.3% for 25-OH vitamin D2. Our results for the DEQAS serum pools averaged −7.2% from the overall LC-MS method mean. The DiaSorin RIA agreed well with the LC-MS/MS method (r2 = 0.90; average bias 1.61 nmol/L), the Roche ECLIA considerably disagreed (r2 = 0.58; bias 10.13 nmol/L). This LC-MS/MS method is reliable and robust for the measurement of both 25-OH vitamin D3 and 25-OH vitamin D2 in human serum.
Keywords :
method comparison , 25-Hydroxyvitamin D , radioimmunoassay , LC-MS/MS
Journal title :
Journal of Chromatography B
Journal title :
Journal of Chromatography B