Title of article :
An optimized and validated RP-HPLC/UV detection method for simultaneous determination of all-trans-Retinol (Vitamin A) and α-Tocopherol (Vitamin E) in human serum: Comparison of different particulate reversed-phase HPLC columns
Author/Authors :
Khan، نويسنده , , Abad and Khan، نويسنده , , Muhammad I. and Iqbal، نويسنده , , Zafar and Shah، نويسنده , , Yasar and Ahmad، نويسنده , , Lateef and Watson، نويسنده , , David G.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2010
Pages :
9
From page :
2339
To page :
2347
Abstract :
A novel, simple and fast reversed-phase HPLC/UV method was developed, optimized for various chromatographic conditions, and validated according to international guidelines for simultaneous determination of all-trans-retinol and α-tocopherol in human serum using retinyl acetate as internal standard in the concentration of 0.5 μg/ml. A liquid-phase extraction was applied to the 250 μl of serum with n-hexane–dichloromethane mixture (70:30, v/v), in two steps, using ethanol–methanol mixture (95:5, v/v) for protein precipitation and BHT (butylated hydroxy toluene) as stabilizer for sample preparation. Both analytes were analyzed on Kromasil 100 C18 column (150 mm × 4.6 mm, 5 μm), Brownlee analytical (Perkin Elmer) C18 column (150 mm × 4.6 mm, 5 μm), and Supelco (Supelcosil) LC-18 column (150 mm × 3 mm, 3 μm), protected by a Perkin Elmer C18 (30 mm × 4.6 mm, 10 μm; Norwalk, USA) pre-column guard cartridge, at 292 nm wavelength, using methanol–water (99:1, v/v), in isocratic mode as mobile phase applied at flow rate of 1.5 ml/min and 1 ml/min for both 5 μm and 3 μm columns, respectively. Complete separation of all the analytes was achieved in 3 and 6 min on 3 μm and 5 μm columns, respectively by injecting 20 μl of sample into the HPLC system by autosampler, keeping column oven temperature at 25 °C. Different particulate reversed-phase chromatographic columns were evaluated in order to select the best column in terms of sensitivity, selectivity, resolution and short run time of both the analytes and it was concluded that 3 μm columns are better to be used in clinical set up as well as in laboratories for the separation of these analytes in a shorter time as compared with 5 μm columns. The method was validated and applied for the analysis of all-trans-retinol and α-tocopherol in the serum of human volunteers.
Keywords :
?-Tocopherol , Validation , Isocratic mode , Chromatographic columns , RESOLUTION , All-trans-retinol
Journal title :
Journal of Chromatography B
Serial Year :
2010
Journal title :
Journal of Chromatography B
Record number :
1468734
Link To Document :
بازگشت