Title of article
Validation of a method for the determination of the anticancer agent Combretastatin A1 phosphate (CA1P, OXi4503) in human plasma by HPLC with post-column photolysis and fluorescence detection
Author/Authors
Stratford ، نويسنده , , Michael R.L. and Folkes، نويسنده , , Lisa K.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2011
Pages
4
From page
2673
To page
2676
Abstract
A validated method for the determination of Combretastatin A1 phosphate (CA1P, OXi4503), a bisphosphate prodrug of the vascular disrupting agent Combretastatin A1 in human plasma has been developed using fluorescence detection after post-column photolysis. The separation used the ion-pairing agent tetrabutylammonium hydrogen sulphate, and this agent was also required to give consistently high recovery from plasma. Initially, the range was shown to be linear (r2 > 0.995) from the LOQ of 0.025 μM to 5 μM, but as the trial progressed to much higher doses, using a lower injection volume, the assay was subsequently subject to limited revalidation to cover the range from 0.05 to 50 μM. Intra-assay precision and accuracy ranged from 2.2 to 11.8% and 1.8 to 13% respectively, and for inter-assay from 4.4 to 14.9% and 1.7 to 6.5%. Mean recovery of OXi4503 from plasma was 80.2%.
Keywords
OXi4503 , PLASMA , HPLC , Ion-pairing , fluorescence , CA1P