Title of article :
Influence of ionization source design on matrix effects during LC–ESI-MS/MS analysis
Author/Authors :
Ghosh، نويسنده , , Chinmoy and Shinde، نويسنده , , Chandrakant P. and Chakraborty، نويسنده , , Bhaswat S.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2012
Pages :
8
From page :
193
To page :
200
Abstract :
Glycerophosphocholines (GPChos) are known to cause matrix ionization effects during the analysis of biological samples (i.e. plasma, urine, etc.) in LC–MS/MS. In general, such matrix effect is directly related to an insufficient sample clean-up of the biofluids. In addition to GPCho; design of ionization source and/or LC also plays a very important role in matrix effects. In this research paper, different types of matrix effects, i.e. ion suppression or enhancement were observed in differently designed ion sources coupled with different LCs, from the same molecule, acamprosate (ACM), under the same chromatographic conditions. ACM was analyzed in a negative polarity in electrospray ionization interface using Z-spray and orthogonal spray ion source design. The analyte showed almost complete ion suppression in the Z-spray ionization source coupled with UPLC/HPLC, whereas there was very little ion enhancement in the orthogonal spray ionization source coupled with HPLC. In both the cases different GPChos were responsible, as evident from the presence of m/z 815.4 in Z-spray ion source and m/z 759.0 in orthogonal spray ion source. Hence, this approach can be used to evaluate the matrix effects in plasma samples during development and validation of LC–MS/MS method of drugs and their metabolites in different biological matrixes.
Keywords :
LC–MS/MS , Matrix effect , Acamprosate , Phospholipids , Orthogonal spray , Z-Spray , Ion source design
Journal title :
Journal of Chromatography B
Serial Year :
2012
Journal title :
Journal of Chromatography B
Record number :
1469860
Link To Document :
بازگشت