Abstract
A fast and sensitive liquid chromatographic (fast-LC) method with radiometric detection was developed and validated to analyze positron emission tomography (PET) radioligands in plasma during PET studies. The plasma samples were deproteinized with acetonitrile and the extracts were injected into the fast-LC system coupled to an on-line radioactivity detector. Under the optimum conditions, complete separation of target PET radioligands from their radioactive metabolites was achieved within the short run time of only 3.5-min. The limits of detection were 1.0–1.2 Becquerel (Bq) for 11C and 18F-labeled compounds. This method can successfully be applied to study the metabolism of a wide variety of PET radioligands in human and monkey plasma with higher numbers of samples to be analyzed compared to the traditional LC method.
Keywords
Radioactive metabolite , Positron emission tomography (PET) , Fast liquid chromatography (fast-LC) , Metabolite analysis , Short-lived radioligand