Title of article :
Pressurized liquid extraction coupled with countercurrent chromatography for systematic isolation of chemical constituents by preprogrammed automatic control
Author/Authors :
Zhang، نويسنده , , Yuchi and Guo، نويسنده , , Liping and Liu، نويسنده , , Chunming and Fu، نويسنده , , Zi?ao and Cong، نويسنده , , Lei S. Qi، نويسنده , , Yanjuan and Li، نويسنده , , Dongping and Li، نويسنده , , Sainan and Wang، نويسنده , , Jing، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2013
Pages :
10
From page :
16
To page :
25
Abstract :
Pressurized liquid extraction (PLE) coupled with high-speed countercurrent chromatography (HSCCC) via an automated procedure was firstly developed to extract and isolate ginsenosides from Panax quinquefolium. The experiments were designed under the guidance of mathematical model. The partition coefficient (K) values of the target compounds and resolutions of peak profiles were employed as the research indicators, and exponential function and binomial formulas were used to optimizing the solvent systems and flow rates of the mobile phases in a three-stage separation. In the first stage, ethyl acetate, n-butanol, and water were simultaneously pumped into the solvent separator at the flow rates 11.0, 10.0, and 23.0 mL/min, respectively. The upper phase of the solvent system in the solvent separator was used as both the PLE solvent and the HSCCC stationary phase, followed by elution with the lower phase of the corresponding solvent system to separate the common ginsenosides. In the second and third stages, rare ginsenosides were first separated by elution with ethyl acetate, n-butanol, methanol, and water (flow rates: 20.0, 3.0, 5.0, and 11.0 mL/min, respectively), then with n-heptane, n-butanol, methanol, and water (flow rates: 17.5, 6.0, 5.0, and 22.5 mL/min, respectively). Nine target compounds, with purities exceeding 95.0%, and three non-target compounds, with purities above 84.48%, were successfully separated at the semipreparative scale in 450 min. The separation results prove that the PLE/HSCCC parameters calculated via mathematical model and formulas were accurately and scientifically. This research has opened up great prospects for industrial automation application.
Keywords :
Panax quinquefolium , High speed countercurrent chromatography , Pressurized liquid extraction , Preprogrammed automatic control
Journal title :
Journal of Chromatography B
Serial Year :
2013
Journal title :
Journal of Chromatography B
Record number :
1471430
Link To Document :
بازگشت