Title of article :
MCM-41 solid phase membrane tip extraction combined with liquid chromatography for the determination of non-steroidal anti-inflammatory drugs in human urine
Author/Authors :
Kamaruzaman، نويسنده , , Sazlinda and Sanagi، نويسنده , , Mohd Marsin and Endud، نويسنده , , Salasiah and Wan Ibrahim، نويسنده , , Wan Aini and Yahaya، نويسنده , , Noorfatimah، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2013
Abstract :
Mesoporous silica material, MCM-41, was utilized for the first time as an adsorbent in solid phase membrane tip extraction (SPMTE) of non-steroidal anti-inflammatory drugs (NSAIDs) in urine prior to high performance liquid chromatography-ultraviolet (HPLC-UV) analysis. The prepared MCM-41 material was enclosed in a polypropylene membrane tip and used as an adsorbent in SPMTE. Four NSAIDs namely ketoprofen, diclofenac, mefenamic acid and naproxen were selected as model analytes. Several important parameters, such as conditioning solvent, sample pH, salting-out effect, sample volume, extraction time, desorption solvent and desorption time were optimized. Under the optimum extraction conditions, the MCM-41-SPMTE method showed good linearity in the range of 0.01–10 μg/mL with excellent correlation coefficients (r = 0.9977–0.9995), acceptable RSDs (0.4–9.4%, n = 3), good limits of detection (5.7–10.6 μg/L) and relative recoveries (81.4–108.1%). The developed method showed a good tolerance to biological sample matrices.
Keywords :
Non-steroidal anti-inflammatory Drugs , urine , mesoporous silica , Solid phase membrane tip extraction
Journal title :
Journal of Chromatography B
Journal title :
Journal of Chromatography B