Title of article
Chemometric modelling for process analyzers using just a single calibration sample
Author/Authors
Oliveira، نويسنده , , F.R.P. and Goldberg، نويسنده , , K. and Liese، نويسنده , , A. and Hitzmann، نويسنده , , B.، نويسنده ,
Issue Information
دوفصلنامه با شماره پیاپی سال 2008
Pages
5
From page
118
To page
122
Abstract
A new calibration method for process analyzers, which requires a chemometric model, is presented. Instead of a huge amount of calibration data, which are usually necessary to calculate a chemometric model, just a single off-line sample is necessary for the proposed procedure. The further required information to calculate the chemometric model is made available in a form of a theoretical process model, whose parameter values will be estimated inherently during the calibration procedure. The functioning of the new procedure is demonstrated by its application to 2D-fluorescence data for the monitoring of a biotransformation process. In the process acetophenone is converted to (S)-1-phenylethanol.
the new calibration method the relative errors of prediction of the substrate and product concentrations are 4.9% and 4.6%, respectively.
Keywords
PLS , Calibration procedure , Process analyzer , Chemometric modeling
Journal title
Chemometrics and Intelligent Laboratory Systems
Serial Year
2008
Journal title
Chemometrics and Intelligent Laboratory Systems
Record number
1489368
Link To Document