Title of article :
Sensitive determination of dissolved sulfide in estuarine water by solid-phase extraction and high-performance liquid chromatography of methylene blue
Author/Authors :
Tang، نويسنده , , Degui and Santschi، نويسنده , , Peter H، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
5
From page :
305
To page :
309
Abstract :
A sensitive method involving solid-phase extraction and HPLC analysis of methylene blue has been developed to measure nanomolar levels of dissolved sulfide in oxic surface waters. The procedure included 1) a preconcentration step, in which methylene blue generated from sulfide reaction with n,n-dimethyl-p-phenylenediamine in acidic conditions in the presence of ferric ion was absorbed onto Waters tC18 cartridge; and 2) a determination step, in which methylene blue was separated by HPLC in a gradient elution to minimize natural organic matter interference and detected by absorbance. The concentrations of the dissolved sulfide, quantified by standard addition, were about 2.1–4.7 nM in oxic surface waters from Galveston Bay, Texas.
Keywords :
Sulfide , Methylene blue
Journal title :
Journal of Chromatography A
Serial Year :
2000
Journal title :
Journal of Chromatography A
Record number :
1501756
Link To Document :
بازگشت