• Title of article

    Comparison of two extraction methods independently developed on two conceptually different automated supercritical fluid extraction systems for the determination of polychlorinated biphenyls in sediments

  • Author/Authors

    Nilsson، نويسنده , , Tobias and Bjِrklund، نويسنده , , Erland and Bّwadt، نويسنده , , Sّren، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    5
  • From page
    195
  • To page
    199
  • Abstract
    Two extraction methods that independently have been developed on conceptually different automated supercritical fluid extraction systems, ISCO SFX 3560 (syringe pump and liquid trapping) and Hewlett-Packard 7680T SFE (reciprocating pump and solid-phase trapping), were compared for the extraction of polychlorinated biphenyls from two Swedish sediments. The results demonstrated that the high-temperature ISCO method in some cases yields a more exhaustive extraction, but also less clean extracts due to co-extraction of unwanted matrix components which are all present in the trapping solvent. The medium-temperature Hewlett-Packard method may sometimes cause problems with quantitative recoveries, but on the other hand it yields very clean extracts due to the extra selectivity resulting from collection on a solid-phase trap.
  • Keywords
    Polychlorinated biphenyls
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2000
  • Journal title
    Journal of Chromatography A
  • Record number

    1502427