Title of article :
Simplified determining procedure for routine residue monitoring of sulphamethazine and sulphadimethoxine in milk
Author/Authors :
Furusawa، نويسنده , , Naoto، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Abstract :
A simplified determining/identifying method for residual sulphamethazine (SMZ) and sulphadimethoxine (SDM) in milk by using a high-performance liquid chromatography (HPLC) with a photo-diode array detector was presented. Both sulphonamides in cow’s milk samples were extracted by only stirring with ethanol followed by an Ultrafree®–MC/Biomax as a centrifugal ultra-filtration unit. For determination/identification of SMZ and SDM, a Mightysil® RP-18 GP Aqua column and a mobile phase of 25% (v/v) ethanol solution (in water) with a photo-diode array detector was used. Average recoveries from spiked SMZ and SDM (10–1000 ng/ml each drug) were ⩾83% with the relative standard deviations between 1.4 and 3.7%. The limit of quantitation (LOQ) were calculated to be 5 ng/ml for SMZ and 10 ng/ml for SDM, respectively. The values were below the MRL/tolerance (SMZ, 25 ng/ml; SDM, 10 ng/ml). The total time and solvent required for the analysis of one sample were <35 min and <2 ml of only ethanol, respectively. No toxic solvents were used. The developed procedure was harmless to the human and environment.
Keywords :
Sulphamethazine , Sulphadimethoxine
Journal title :
Journal of Chromatography A