  • Abstract
    This paper deals with the development of practical approaches to a new liquid-delivery system for capillary liquid chromatography. Under different chromatographic conditions, the factors affecting liquid-delivery performance are theoretically described, and the new liquid-delivery system without any splitter is evaluated with its flow-rate accuracy and precision using a variety of solvents. The experimental results demonstrate that the liquid-delivery system is capable of generating accurate, reproducible and conditions-independent micro- and nano-flows.