Title of article
New micro-flow pumping system for liquid chromatography
Author/Authors
Zhou، نويسنده , , X and Furushima، نويسنده , , N and Terashima، نويسنده , , C and Tanaka، نويسنده , , H and Kurano، نويسنده , , M، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2001
Pages
7
From page
165
To page
171
Abstract
This paper deals with the development of practical approaches to a new liquid-delivery system for capillary liquid chromatography. Under different chromatographic conditions, the factors affecting liquid-delivery performance are theoretically described, and the new liquid-delivery system without any splitter is evaluated with its flow-rate accuracy and precision using a variety of solvents. The experimental results demonstrate that the liquid-delivery system is capable of generating accurate, reproducible and conditions-independent micro- and nano-flows.