• Title of article

    Influence of buffer zone concentrations on efficiency in partial filling micellar electrokinetic chromatography

  • Author/Authors

    Michalke، نويسنده , , Daniela and Kolb، نويسنده , , Stephan and Welsch، نويسنده , , Thomas، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    10
  • From page
    113
  • To page
    122
  • Abstract
    The potential of counter pressure-moderated partial filling micellar electrokinetic chromatography (PF-MEKC) was investigated in this work. Plate numbers of homologous ω-phenylalcohols were measured in a two-plug PF-MEKC system varying the concentrations and hence the ionic strengths of the background buffer compared to the sodium dodecyl sulfate-containing separation buffer and the counter pressure on the cathodic buffer reservoir. It was observed that plate numbers are strongly influenced by both the buffer concentrations and the counter pressure. Highest plate numbers were obtained with a buffer system where the concentrations are adjusted such that the electroosmotic flow velocities in both zones are equal. Differences in the local electroosmotic flow velocities of the zones caused by different buffer concentrations are responsible for tremendously reduced plate numbers. The efficiency drop is explained in several models by the formation of an intersegmental pressure which produces a parabolically shaped laminar flow component in both zones. Thus, the electroosmotic plug-like flow profile is distorted and the efficiency is reduced. The effect of counter pressure on efficiency turned out to be very complex in dependence on the buffer system applied.
  • Keywords
    Phenylalcohols
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2001
  • Journal title
    Journal of Chromatography A
  • Record number

    1507816