Title of article :
Determination of heparin on intraocular lens surfaces by ion chromatography
Author/Authors :
Ander، نويسنده , , Birgitta and Karlsson، نويسنده , , Anders and ضhrlund، نويسنده , , إke، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
6
From page :
105
To page :
110
Abstract :
A sensitive and selective method has been developed for the determination of heparin on heparin coated PMMA, poly(methyl methacrylate), intraocular lenses. Heparin was hydrolysed to glucosamine and glucuronic acid, and the content of glucosamine was determined using ion chromatography with pulsed amperometric detection. In order to verify that a complete hydrolysis was obtained for the heparin on the coated intraocular lenses, electron spectroscopy for chemical analysis (ESCA) was used for analysing traces of sulphur on the lens surfaces. The sensitivity of the method allows quantitative determination of 150 ng of heparin on one individual lens. The new method was compared to a standard spectrophotometric method, measuring the colour intensity of a heparin toluidine blue complex. Correlation between the methods was shown for samples prepared from PMMA lenses coated with different amounts of heparin.
Keywords :
HEPARIN
Journal title :
Journal of Chromatography A
Serial Year :
2001
Journal title :
Journal of Chromatography A
Record number :
1507888
Link To Document :
بازگشت